X-ray Tube

X-ray Tube Assemblies for Angiography Systems

Below information is purpose for of development, design and production of medical device systems by implementing our company's products. Technical and regulatory confirmation/verification is necessary before purchasing and using of these products.

LM-Angiography x-ray tube

LM-Angiography x-ray tube

Features

  • Continuous high-speed rotation is made by the use of hydrodynamic liquid metal bearings.
  • These tubes employ a grid control function that enables high-speed pulsed fluoroscopy, which is used in high speed applications such as cine fluoroscopy.

Lineup

Click photo to enlarge.

X-ray Tube Assemblies for Angiography Systems Lineup Table
Image Model
Name
Focal
Spot
L/M/S
Max.
Rating
(0.1s)
(kW)
Long
-time
Maxi
mum
Input
(W)
Max. Voltage
(kV)
Max.
Tube
Current
(mA)
Anode Housing Assembly Down
load
Angle
(°)
Dia
meter
(mm)
Max.
Heat
Content
Cooling
Rate
(W)
Max.
Rotation
Speed
(min-1)
Max.
Heat
Dissi
pation
(W)
Housing
Type
Weight*
(kg)
R F
Conti
nuous
Grid
Control
(kJ) (kHU)
E79016X(L) E79016X L 1.0 100 2,200 125 125 120 1,100 11 140 2,130 3,000 5,500 9,000 3,500 XH-177 46 Links with this mark will open in a separate window.
(328KB)
M 0.6 48 650
S 0.3 17 180
E7902X(L) E7902X L 0.8 90 2,200 125 125 110 950 8 140 2,130 3,000 5,500 9,000 3,500 XH-177 46 Links with this mark will open in a separate window.
(334KB)
M 0.5 45 600
S
E79036X(L) E79036X L 0.8 80 2,200 125 125 120 800 8 120 1,500 2,100 4,500 9,000 3,000 XH-1008 38 Links with this mark will open in a separate window.
(332KB)
S 0.5 44 470
E79039X(L) E79039X L 1.0 80 2,200 125 125 120 800 11 120 1,500 2,100 4,500 9,000 3,000 XH-1008 38 Links with this mark will open in a separate window.
(313KB)
S 0.6 43 480
Notes
* Without heat exchanger.