For Mobile Surgical X-ray System (Mobile C-arm)

Below information is purpose for of development, design and production of medical device systems by implementing our company's products.
Technical and regulatory confirmation/verification is necessary before purchasing and using of these products.

Lineup

Stationary Anode X-ray Tubes Medical X-ray Tubes

Click photo to enlarge.

Stationary Anode X-ray Tubes Medical X-ray Tubes 
Image Type
(Series)
Focal
Spot
Max.
Rating
(1 s)
(W)
Max.
Voltage
(kV)
Max.
Current
(mA)
Circuit Target
Angle
(degree)
Anode Provide Down
load
without S SB
Heat
Capacity
(kJ)
Cooling
Rate
(W)
Dimensions(mm)
Length×Diameter
D-063R D-063R 0.6 2,020 110 20 C 8 32 600 141×52 151×60 - -
DF-151R DF-
151R
0.5/
1.5
680/
3,200
110 15/60 C 16 35.5 600 160×52 166×60 166×64 Links with this mark will open in a separate window.
(326KB)
Notes
Circuit
C : Constant Potential High-Voltage Generator(All tubes are center grounded)
SF : Two-peak high-voltage generator
AC : One-Peak High-Voltage Generator(Self-rectified)

X-ray Flat Panel Detectors 

Click photo to enlarge.

X-ray Flat Panel Detectors 
Image Inch Type Model Name Useful Entrance
Field Size
Number of
Pixels
Pitch
(μm)
Interface Application Down
load
Dynamic-Type
FDX3334RF 13×13 FDX3334RF 330×343 2304×2400 143 LC-Duplex
Optical Fiber
Fluoroscopy&Radiography
(Static & Dynamic image)
Links with this mark will open in a separate window.
(629KB)
FDX2121F 21×21(cm) FDX2121F 210×210 1024×1024 205 Ethernet
(1000 Base-T)
Mobile C-arm
(Static & Dynamic image)
Links with this mark will open in a separate window.
(316KB)

X-ray Image Intensifiers Dental/Medical 6-inch

Click photo to enlarge.

X-ray Image Intensifiers Dental/Medical 6-inch
Image Model Name Nominal Entrance
Field Size
Useful Entrance
Field Size
Output
Image
Dia-
meter
(mm)
Central Resolution
(Lp/cm) typical
Contrast Ratio
typical
DQE
(IEC Stan-
dard)
Mechanical Appli-
cation
Download
Normal
mode
Normal
mode
M
mode
Nor-
mal
mode
M
mode
10% area 10 mm dia. typical Overall length Mounting Surface
E5863SD-P6A E5863SD-
P6A
150 mm
min.
140 mm min. - 20 ±0.5 54 - 30:1 20:1 65% 276
±5
mm
Side of the
image
intensifier
&
Front of the
image
intensifier
C-arm/
Fluoro
table
-
E5883SD-
P6A
105 ±5 mm 64 -
Notes
Products with a model name ending in "A" are 24 Vdc input type.

X-ray Image Intensifiers Dental/Medical 9-inch

Click photo to enlarge.

X-ray Image Intensifiers Dental/Medical 9-inch
Image Model Name Nominal Entrance Field Size Useful Entrance
Field Size
Output
Image
Dia-
meter
(mm)
Central Resolution
(Lp/cm) typical
Contrast Ratio
typical
DQE
(IEC Stan-
dard)
Mechanical Appli-
cation
Download
Normal
mode
Normal
mode
M1
mode
M2
mode
Nor-
mal
mode
M1
mode
M2
mode
10% area 10 mm dia. typical Overall length Mounting Surface
E5764SD-P4A E5764SD-
P4A
230 mm min. 215 mm min. 160 ±5 mm 120 ±5 mm 20 ±0.5 48 56 66 25:1 16:1 65% 338 ±5
mm
Same side as the power supply box
&
Front of the image intensifier
C-arm/
Fluoro table
-
E5830SD-
P4A
25 ±0.5 52 58 68 30:1 19:1 Links with this mark will open in a separate window.
(291KB)
E5764SD-P7 E5804SD-P7A 160 ±5 mm 120 ±5 mm 20 ±0.5 48 - - 25:1 16:1 341 ±5 mm Same side as the power supply box C-arm Links with this mark will open in a separate window.
(779KB)
E5764SD-P7A 56 66 Links with this mark will open in a separate window.
(705KB)
E5830SD-P7A 25 ±0.5 52 58 68 30:1 19:1 Links with this mark will open in a separate window.
(705KB)
Notes
Products with a model name ending in "A" are 24 Vdc input type.

X-ray Image Intensifiers Dental/Medical 12-inch

Click photo to enlarge.

X-ray Image Intensifiers Dental/Medical 12-inch
Image Model Name Nominal Entrance Field Size Useful Entrance
Field Size
Output
Image
Dia-
meter
(mm)
Central Resolution
(Lp/cm) typical
Contrast Ratio
typical
DQE
(IEC Stan-
dard)
Mechanical Appli-
cation
Download
Normal
mode
Normal
mode
M1
mode
M2
mode
Nor-
mal
mode
M1
mode
M2
mode
10% area 10 mm dia. typical Overall length Mounting Surface
E5796SD-P2A E5796SD-
P2A
310
mm
min.
290 mm min. 215 ±5 mm 160 ±5 mm 25 ±0.5 46 50 56 30:1 18:1 65% 434.5
±5
mm
Side of the image intensifier
&
Front of the image intensifier
C-arm/
Fluoro
table
Links with this mark will open in a separate window.
(275KB)
Notes
Products with a model name ending in "A" are 24 Vdc input type.