X-ray Tube

X-ray Tube Assemblies for CT Systems

Below information is purpose for of development, design and production of medical device systems by implementing our company's products. Technical and regulatory confirmation/verification is necessary before purchasing and using of these products.

An LM-CT tube is a compact ultra-high cooling X-ray tube that uses hydrodynamic bearings in its rotation mechanism and an all-metal (tungsten and molybdenum) high capacity anode.

Hydrodynamic bearings contribute lower vibration, lower noise, and greater reliability in terms of all-metal anode rotation. Temperature distribution uniformity in all-metal anodes enhances radiation cooling.

X-ray Tube Assemblies for CT Systems

Lineup

Click photo to enlarge.

X-ray Tube Assemblies for CT Systems Lineup Table
Image Model
Name
Focal
Spot
L/S
Max.
Rating
(kW)
Long
-time
Maxi
mum
Input
(kW)
X-ray
Tube
Voltage
(kV)
Max.
Tube
Current
(mA)
Anode Max. G
Proof in Gantry Rotation (G)
Housing Assembly Down
load
Angle
(°)
Dia
meter
(mm)
Max.
Heat
Content
Max.
Heat
Dissi
pation (kW)
Max.
Rota
tion
Speed
(min-1)
Max.
Heat
Dissi
pation (kW)
Housing Type Weight*(kg)
Min. Max. (kJ) (kHU)
E79005X(L) E79005X L 1.7×1.7 36 2.0 80 135 300 7 132 1,420 2,000 4.0 2,700 6.3 3.6 XH-168 38 Links with this mark will open in a separate window.
(384KB)
S 1.1×1.3 24 220
- XRC-4551X L 1.4×1.4 48 4.0 - 135 400 7 140 2,840 4,000 10.2 6,360 8.6 4.0 XH-168 41 Links with this mark will open in a separate window.
(401KB)
S 0.9×0.7 30 300
- XRC-4552X L 1.4×1.4 48 4.0 - 135 400 7 140 2,840 4,000 10.2 6,360 12 4.0 XH-168 41 Links with this mark will open in a separate window.
(403KB)
S 0.9×0.7 30 300
Notes
* Without heat exchanger.

The product information shown in the lineup table is not intended as an advertisement or inducement for the Japanese domestic market.